Hot cell for the operation of automated synthesis modules
The mini cell is especially designed for the operation of customary PET synthesis modules or hot cells adapted for research or nuclear medical purposes. The GMP-compliant design ensures the quality of production conditions.
There are four different types, Type I, Type II, Type III and Type IV. The standard model consists of a stainless steel cell and 75 mm lead shielding. A shielded door with an optional lead glass window is implemented at the front side for easy access to the synthesis module.
Customizable design of the mini cell and the thickness of the shielding for customer intended process and / or physical conditions.
Also available are special solutions for the handling of Ga-68 which include coating the inner cell with polyethylene for handling hydrochloric acid.
Options
- Adjustment of the dimensions and the height of the cladding
- Pull- out for synthesis module
- Fittings and bushings for gas pipes
- Double mini cell arranged vertically one above the other
- Laptop holder
- Shielded bushings for nuclides
- Storage box for radioactive waste
- Specially for 68Ga: PE coating of the inner cell for handling Hydrochloric acid
- Shielded generator storage box below the mini cell with sealed bushings to the inner cell
Dimensions / Weight
|
Internal (WxDxH) |
External (WxD) |
Height |
Weight |
Type I | 700x560x775mm | 920x850 mm | 3000 mm | ~3,500 kg |
Type II | 700x640x775mm | 920x930 mm | 3000 mm | ~3,800 kg |
Type III | 850x640x775mm | 1070x930 mm | 3000 mm | ~4,500 kg |
Type IV | 700x500x610mm | 920x770 mm | 3000 mm | ~6,000 kg |
to do:
material | Inner cell stainless steel (1.4301) |
shielding | 75 mm lead on all sides |
door | |
execution | Swing door hinged on the left or right |
Radiation protection window | 145x145 mm (WxH), 75 mm lead equivalent |
Tightness inner cell | Leakage rate <0.25% / h according to DIN 25412-2 |
Ventilation technology | |
Supply air | H13 filter, connection DN 25 |
Exhaust air | H13 filter with activated carbon, connection DN 25 |
Vacuum regulation | manually operated ball valves |
Monitoring | Differential pressure |
Electrics / pneumatics | |
Power connection | 3x230 / 400 V 50 Hz N / PE, 16 A back-up fuse |
consumption | approx. 1 kW |
Sockets | 2 pieces inside the cell (IP54), switchable from the outside, 2 pieces above / below |
Compressed air | the cell min. 6 bar |
Cladding | |
material | powder coated steel plates |
color | Pure white (RAL 9010) or according to customer requirements |